SPECIFICATION 

Parameters

33: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PCT, MPV, PDW, P-LCR, NE, NE%, LY, LY%, MO, MO%, EO, EO%, BA, BA%, P-LCC*, Mentzer Index*, RDWI*, IG%*, IG#*, Band%*, Band#*, Seg%*, Seg#*

(*) RUO parameters (Research Use Only)

Reagents 5: 2 hemolyzing reagents, 1 diluent, 2 detergents
Sample volume

CBC : 25 μL

CBC+DIFF : 40 μL

Pre-dilution mode : 20 μL

Thoughput Up to 90 samples /hour
 

WBC : 2.0% or less (WBC : 4.00 × 103/μL or more)

RBC : 1.5% or less (RBC : 4.00 × 106/μL or more)

HGB : 1.5% or less

HCT : 1.5% or less MCV : 1.0% or less

PLT : 4.0% or less (PLT : 100 × 103/μL or more)

NE% : 5.0% or less (NE% : 30.0% or more AND WBC : 4.00 × 103/μL or more)

LY% : 5.0% or less (LY% : 15.0% or more AND WBC : 4.00 × 103/μL or more)

MO% : 12.0% or less (MO% : 5.0% or more AND WBC : 4.00 × 103/μL or more)

EO% : 20.0% or less OR within ±1.0 EO% (WBC : 4.00 × 103/μL or more)

BA% : 30.0% or less OR within ±1.0 BA% (WBC : 4.00 × 103/μL or more)

NE : 8.0% or less (NE : 1.20 × 103/μL or more)

LY : 8.0% or less (LY : 0.60 × 103/μL or more)

MO : 20.0% or less (MO : 0.20 × 103/μL or more)

EO : 25.0% or less OR within ±0.10 × 103/μL (WBC : 4.00 × 103/μL or more)

BA : 30.0% or less OR within ±0.10 × 103/μL (WBC : 4.00 × 103/μL or more)

(Specifications above applies to normal mode)

Complete QC programs L & J / Westgard Multi rule / X bar B(X bar Batch) management
Options

Software kit MEK-9100, QS-023W

Extra sample racks, T411A

Waste sensor, JW-910W

Dimensions and weight 675 W × 589 D × 576 H mm, 66 kg